The valve can be installed directly downstream of the primary cooler to protect more vulnerable components further down the line. This is advantageous for the protection of personnel and instruments that could otherwise be harmed by high temperature samples. The valve responds only to temperature. Under normal operation the valve is fully open. If the sample temperature reaches 115°F (46°C), the thermal element within the valve begins to close the inlet orifice. The inlet orifice is fully closed when the sample reaches 125°F (52°C). The valve automatically opens as the element cools.
Advantages:
- The valve is automatic, with no reset or operator involvement required
- Features corrosion-resistent stainless steel and EPDM materials to reduce contamination of sample stream
- The valve is compact, and the low mass means fast response – the overall size is 1-3/8-inch diameter by 3-1/2-inch length
- Reliable shut-off with a ram-type plug design for a tight seal upon shut-off