Ammonium Analyzer AM200

Direct ammonium measurement based on UV spectroscopy.

Measurement of ammonium in water and waste water in 3 minutes without filtration.

The AM200 analyser has been developed to match with most of applications where a continuous analysis of NH4+ in water is necessary.

The AM200 uses a UV spectrophotometric measurement method. This very reliable method gives the AM200 the advantage of a very long term stability of the measurements and makes the analyser insensitive to probe damage due to aggressive water.

A very low amount of NaOH (1 mL of diluted NaOH solution) is first injected in sampled water in order to perform the measurement. The chemical equilibrium between NH4+ and NH3 in water is displaced in favour of NH3 because of the pH. This displacement of equilibrium is pretty total if pH>12 and causes NH3 to give off from the solution. A certain amount of NH3 depending on temperature passes in gas phase and UV spectrometric measurement is then performed on the gas.

HENRY’s law related to liquid/gas equilibrium is then applied and allows the precise determination of the amount of ammonium present in water by the analysis of outgased NH3.

Advantages:

  • Online ammonium monitoring: 0,15 – 20 mg/L NH4+; 0,5 – 1000 mg/L NH4+; other ranges on request
  • Quick results : measurement in less than 3 minutes.
  • Insensitive to turbidity and water coloration: Exclusive measurement method on gas phase.
  • Reliable measurement in the long run: Automatic cleaning system; Zero adjustment system.
  • Compact size.
Read more
Weight : 20 Kg
Range : 0.1 – 20 mg/L (clean water)
0 – 1000 mg/L (effluent, waste water…)
Accuracy +/-0.15 mg/L at 0.5 mg/L and +/-0.5 mg/L at 10 mg/L.

The initial calibration accuracy is within 4%

Energy : 100-240V 50/60 Hz 60 W
+ internal battery 12 V
Outputs : 4-20 mA isolated, 12 bits resolution
High threshold/low threshold relay
Alarm default relay
Enclosure : Watertight IP55 Painted steel box for electronics and
spectrophotometer.
Polyester IP65 enclosure for hydraulic system.
Communication : RS232 or RS485 (on demand) for PC, or MODBUS
Water flow rate : typical 0.6 L/mn
Measurement cell volume : 50 mL
Sample temperature : > 3 to 40°C (in all the cases the temperature of
measured water must be lower than that of the air
surrounding the analyser).
Sample pressure : No pressure if sampling pump
0.1 to 1 bar without sampling pump