Silica analyzer – PCA200-SiO2

Online silica analyzer. Colorimetric measurement of silica in 5 minutes on any type of water, without filtration.

The reliable control of several critical parameters is required in many industrial processes. Also environmental regulations become a more and more important issue in waste water treatment and discharge. For a large number of inorganic as well as organic substances discharge limits are introduced and enforced by environmental authorities. To be able to satisfy the growing demand for reliable automatic analytical instruments DATALINK INSTRUMENTS developed the Process Colorimetric Analyser (PCA200). The PCA200 is designed as a very flexible instrument that can be customized easily for almost any colorimetric or photometric analysis.

Due to its flexible design the instrument can be adapted to any colorimetric method.

The technology used for this measurement is the colorimetric method. This method is performed in most cases without water filtration thanks to turbidity compensation treatment during the measurement.

The PCA200 offers the possibility to compensate turbidity effects. There are two options depending on the application:

  • Running a blank on the sample during each measurement, or

Measuring the absorbance at a second wavelength and compensate the influence mathematically.

Advantages:

  • Online silica measurement: (0-1mg/L to 0-20mg/L)
  • Colorimetric method for detecting very low concentrations
  • Quick results : Measurement in 5 minutes
  • Reliable long-term measurement: Automated cleaning system and zero adjustment; Automatic calibration from a standard.
  • Functional measurement of dirty water: The technology at several wavelengths allows the measurement of the dirty water without filtration; Large section hoses limit risk of clogging.
Read more
Analytical method Colorimetric method
Range 0-1mg/L to 0-20 mg/L SiO2
Accuracy +/- 5% of full scale
Minimum measuring interval 5 – 30 min. Application dependent
Sample pressure Up to 4 bars with sampling valve

< 0.2 bar with sampling pump

Sample flow rate Typical 0.2 to 0.5 L/mn
Sample temperature 0.5 – 60 °C (sample cooler available as an option)
Optical system 2-Wavelengths photometer
Sample input/output Input:

w/ sampling pump : Fast connectors for silicone tube 6.4 mm inner diameter.

w/ sampling valve : Double ring connector for 12 mm external tube diameter

Output :

Fast connectors for silicone tube 6.4 mm inner diameter

Power supply 110-120 V/ 220-240V 50/60 Hz 30 VA

Internal battery 12 V

Outputs Four 4-20 mA insulated outputs, 12 bits resolution

High/low level alarm relays (potential free, NC/NO)

Analyser fault alarm relay (potential free, NC/NO)

Mains presence relay (potential free, NC/NO)

Communication RS232/485 for PC, modem or MODBUS
Environmental conditions Rel. Humidity : 5 – 95% (not condensing)

Temperature : 10 – 40 °C

Housing: Watertight IP66 Painted steel box
Weight +/- 30 Kg
Read more
Video

Contact info product